+86-21-54286005 info@tenmed.net Room 602, Building 1, No. 111 Luxiang Road (Greenland Park Plaza), Baoshan District, Shanghai, China
Language: English

Mindray BC-6800 Plus Auto Hematology Analyzer

Mindray BC-6800 Plus Auto Hematology Analyzer

Phone:+86-21-54286005

Resources

BC-6800_plus_Hematology_Analyzer_Brochure

Auto Hematology Analyzer

BC-6800Plus

More than Fast

 

More than Fast

The advanced BC-6800Plus not only provides more reliable routine CBC plus 5-part DIFF results, but also has a higher throughput of up to 200 samples per hour. Based on the new SF Cube technology platform, the BC-6800Plus could provide more reliable blood cell results and abnormal cells flagging results with anticipated warning messages, exceeding clinicians’ expectations and setting a new standard for blood analysis.

 

 

SF Cube Technology

With the newly designed opticals and reagent systems, the SF Cube technology can help doctors to better differentiate the clusters of cells, which is the key to revealing more abnormal cells.

SF Cube Technology

 

NRBC result in every CBC

In WNB scattergram, BC-6800Plus provides NRBC, Basophils and WBC-N* results. It means that the actual number of NRBCs can be measured in routine CBC, if they are present in the sample.

Basophils are counted in this counting channel with NRBC results.

Basophil and NRBC results are generated on BC-6800Plus without extra reagent or cost.

NRBC result in every CBC

 

NLR parameter in each CBC+DIFF

  • Neutrophil-to-lymphocyte ratio (NLR) is calculated by dividing neutrophil count by lymphocyte count, usually from peripheral blood sample.
  • In laboratory medicine, NLR is used as a marker of subclinical inflammation.
  • According to clinical studies, NLR can be used as a prognostic factor of severe illness in COVID-19 patients. 

NLR parameter in each CBC+DIFF

 

More accurate RET and PLT-O result

By utilizing the new fluorescent dye, the reticulocytes and platelets are more specific stained with stronger fluorescent signal, that brings more reliable reticulocyte and platelet results.

 

Automatic Rerun & Reflex

Should the sample results trigger the criteria, the sample rack can automatically step back for a “rerun” or “reflex” check.

Automatic Rerun & Reflex

 

200 test/hour

BC-6800Plus can load up to 100 samples at a time and offers a throughput of up to 200 tests per hour.

200 test/hour

Principles
SF Cube* method to count WBC, 5-part diff, NRBC, RET and PLT-O
Focusing Flow-DC method for RBC and PLT
Cyanide free reagent for hemoglobin test
*S: Scatter; F: Fluorescence; Cube: 3D analysis

Parameters
37 Reportable parameters (whole blood): WBC, Lym%, Mon%,Neu%, Bas%, Eos%, IMG%, Lym#, Mon#, Neu#, Eos#, Bas#, IMG#;RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, NRBC#,NRBC%; PLT, MPV, PDW, PCT, P-LCR, P-LCC, RET%, RET#, RHE, IRF,LFR, MFR, HFR, IPF
42 Research parameters (whole blood): HFC#, HFC%, RBC-O, PLT-O,PLT-I, WBC-O, WBC-D, TNC-D, IME%, IME#, H-NR%, L-NR%, NLR, PLR,WBC-N, TNC-N, InR#, InR‰, Micro#, Micro%, Macro#, Macro%, RPI,H-IPF, IPF#, MRV, FRC#, FRC%, PDW-SD, Neu-X, Neu-Y, Neu-Z,Lym-X, Lym-Y, Lym-Z, Mon-X, Mon-Y, Mon-Z, SRBC, LRBC, SMCV,LMCV(Source: tenmed.net)
7 Reportable parameters (body fluid): WBC-BF, TC-BF#, MN#, MN%,PMN#, PMN%, RBC-BF
11 Research parameters (body fluid): Eos-BF#, Eos-BF%, Neu-BF#,Neu-BF%, HF-BF#, HF-BF%, RBC-BF, LY-BF#, LY-BF%, MO-BF#,MO-BF%

2 Histograms for RBC and PLT 
3 Three-dimension scatter grams: DIFF, WNB, RET 
5 Two-dimension scatter grams: DIFF, WNB, RET, RET-EXT, PLT-O 

Mode
CBC, CBC+DIFF, CBC+DIFF+RET, CBC+RET, RET

Data storage capacity
Up to 10,0000 results including numeric and graphical information

Operating environment
Temperature: 15℃~32℃
Humidity:30%~85%

 

Performance

Parameter

Linearity Range

Precision

Carryover

WBC

0-500×109/L

≤2.5% (≥4×109/L)

≤1.0%

RBC

0-8.60×1012/L

≤1.5% (≥3.5×1012/L)

≤1.0%

HGB

0-260g/L

≤1.0% (110-180g/L)

≤1.0%

HCT

0-75%

≤1.5% (30%-50%)

≤1.0%

PLT

0-5000×109/L

≤4.0% (≥100×109/L)

≤1.0%

RET#

0-0.8×1012/L

≤15% (RBC≥3×1012/L;

1%_RET%_4%)

/

 

Sample volume
Whole blood (Autoloader)   190uL
Whole blood(Open vial) 130uL
Predilute(Open vial) 40uL
Body fluid (Open vial) 130uL


Throughput
Up to 200 samples per hour (CBC+DIFF)
Up to 120 samples per hour (CBC+DIFF+RET)
Up to 120 samples per hour (RET)
Up to 40 samples per hour (Body fluid)


Loading capacity
Up to 100 sample tubes

 



选择您的地区和语言