Mindray BC-6800 Plus Auto Hematology Analyzer
Mindray BC-6800 Plus Auto Hematology Analyzer
Phone:+86-21-54286005
Resources

Microsystem
Endoscopysystem
Energysystem
Mindray BC-6800 Plus Auto Hematology Analyzer
Phone:+86-21-54286005
Auto Hematology Analyzer
More than Fast
More than Fast
The advanced BC-6800Plus not only provides more reliable routine CBC plus 5-part DIFF results, but also has a higher throughput of up to 200 samples per hour. Based on the new SF Cube technology platform, the BC-6800Plus could provide more reliable blood cell results and abnormal cells flagging results with anticipated warning messages, exceeding clinicians’ expectations and setting a new standard for blood analysis.
SF Cube Technology
With the newly designed opticals and reagent systems, the SF Cube technology can help doctors to better differentiate the clusters of cells, which is the key to revealing more abnormal cells.
NRBC result in every CBC
In WNB scattergram, BC-6800Plus provides NRBC, Basophils and WBC-N* results. It means that the actual number of NRBCs can be measured in routine CBC, if they are present in the sample.
Basophils are counted in this counting channel with NRBC results.
Basophil and NRBC results are generated on BC-6800Plus without extra reagent or cost.
NLR parameter in each CBC+DIFF
More accurate RET and PLT-O result
By utilizing the new fluorescent dye, the reticulocytes and platelets are more specific stained with stronger fluorescent signal, that brings more reliable reticulocyte and platelet results.
Automatic Rerun & Reflex
Should the sample results trigger the criteria, the sample rack can automatically step back for a “rerun” or “reflex” check.
200 test/hour
BC-6800Plus can load up to 100 samples at a time and offers a throughput of up to 200 tests per hour.
Principles
SF Cube* method to count WBC, 5-part diff, NRBC, RET and PLT-O
Focusing Flow-DC method for RBC and PLT
Cyanide free reagent for hemoglobin test
*S: Scatter; F: Fluorescence; Cube: 3D analysis
Parameters
37 Reportable parameters (whole blood): WBC, Lym%, Mon%,Neu%, Bas%, Eos%, IMG%, Lym#, Mon#, Neu#, Eos#, Bas#, IMG#;RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, NRBC#,NRBC%; PLT, MPV, PDW, PCT, P-LCR, P-LCC, RET%, RET#, RHE, IRF,LFR, MFR, HFR, IPF
42 Research parameters (whole blood): HFC#, HFC%, RBC-O, PLT-O,PLT-I, WBC-O, WBC-D, TNC-D, IME%, IME#, H-NR%, L-NR%, NLR, PLR,WBC-N, TNC-N, InR#, InR‰, Micro#, Micro%, Macro#, Macro%, RPI,H-IPF, IPF#, MRV, FRC#, FRC%, PDW-SD, Neu-X, Neu-Y, Neu-Z,Lym-X, Lym-Y, Lym-Z, Mon-X, Mon-Y, Mon-Z, SRBC, LRBC, SMCV,LMCV(Source: tenmed.net)
7 Reportable parameters (body fluid): WBC-BF, TC-BF#, MN#, MN%,PMN#, PMN%, RBC-BF
11 Research parameters (body fluid): Eos-BF#, Eos-BF%, Neu-BF#,Neu-BF%, HF-BF#, HF-BF%, RBC-BF, LY-BF#, LY-BF%, MO-BF#,MO-BF%
2 Histograms for RBC and PLT
3 Three-dimension scatter grams: DIFF, WNB, RET
5 Two-dimension scatter grams: DIFF, WNB, RET, RET-EXT, PLT-O
Mode
CBC, CBC+DIFF, CBC+DIFF+RET, CBC+RET, RET
Data storage capacity
Up to 10,0000 results including numeric and graphical information
Operating environment
Temperature: 15℃~32℃
Humidity:30%~85%
Performance
Parameter |
Linearity Range |
Precision |
Carryover |
WBC |
0-500×109/L |
≤2.5% (≥4×109/L) |
≤1.0% |
RBC |
0-8.60×1012/L |
≤1.5% (≥3.5×1012/L) |
≤1.0% |
HGB |
0-260g/L |
≤1.0% (110-180g/L) |
≤1.0% |
HCT |
0-75% |
≤1.5% (30%-50%) |
≤1.0% |
PLT |
0-5000×109/L |
≤4.0% (≥100×109/L) |
≤1.0% |
RET# |
0-0.8×1012/L |
≤15% (RBC≥3×1012/L; 1%_RET%_4%) |
/ |
Sample volume
Whole blood (Autoloader) 190uL
Whole blood(Open vial) 130uL
Predilute(Open vial) 40uL
Body fluid (Open vial) 130uL
Throughput
Up to 200 samples per hour (CBC+DIFF)
Up to 120 samples per hour (CBC+DIFF+RET)
Up to 120 samples per hour (RET)
Up to 40 samples per hour (Body fluid)
Loading capacity
Up to 100 sample tubes