Mindray BC-5390 Auto Hematology Analyzer
BC-5390 Auto Hematology Analyzer
Phone:+86-21-54286005
Microsystem
Endoscopysystem
Energysystem
BC-5390 Auto Hematology Analyzer
Phone:+86-21-54286005
The new BC-5390 Auto Hematology Analyzer provides rapid and reliable test from just 33uL of blood. Utilizing three
mainstream technologies: laser scatter, fiow cytometry and chemical dye, BC-5390 can provide accurate difierential readout.
In order to save time and increase walk-away automation, an autoloader is equipped to hold 40 tubes one time and achieve 60 samples/hr throughput. With the popular windows based software, you can easily perform the routine tests, manage patient results, set up auto-cleaning and connect with LIS server. BC-5390 is your ideal choice to streamline daily workfiow.
The semi-conductor laser system collects forward scatter and side scatter information of white blood cells in a fiow cell and difierentiates lymphocytes,monocytes, neutrophils and eosinophils according to the cell size and granule complexity. Specific chemical dye treatment to eosinophils can separate from neutrophils in DIFF scattergram even on Eosinophilia sample . In addition,basophils are measured in an impedancechannel upon lysing action on the RBC and other WBC groups.
Eosinophilia sample
Mindray patented digital sheath fiow and weak signal detection techniques are applied to ensure accurate counting and sizing of RBC and PLT in the impedance channel. Dynamic discriminator adjustment can improve the separation of RBC and PLT groups when large platelet and micro RBC are present.
Benefiting from these improvements,MCV, RDW and PLT performance are greatly enhanced.
Only 4 routine M-5 reagents are utilized in counting cycle featuring economical consumption and 2 years shelf life. To help precision monitoring, three levels of BC-5D controls are offered in a ready-to-use kit and the assay value table can be automatically imported through USB memory.
The analyzer's windows-based software is simple to use and plays a powerful information hub to store 100000 patient results. You can set-up password access, reference range,auto-cleaning schedule and so on. Also, the built-in report format tool can help to customize the final report type to include microscopic counting, ESR, blood type and
diagnostic remarks.
60 QC files are designed to store L-J QC results. 300 data points can be recorded in single file. 2 common QC programs are enabled for full quality assurance purposes.Patient data is archived and can be searched and presented in trend curve for case follow-up.
Difierent holder adapters supports three types of sample including
Venous whole blood sample, Predilute sample and Capillary whole blood sample. Capillary whole blood sample can be well tested directly which is more convenient for users and makes the analyzers an ideal choice to work with pediatric and geriatric samples.
Venous whole blood sample Pre-diluted sample Capillary whole blood sample
Anti-aging samples with more than 12 hours transportation to central
labs which is a quite frequent and normal situation for small to medium labs in countryside. Hence, diagnosing anti-aging samples accurately becomes an significant issue for developing countries or cities. With enhanced reagent system and upgraded reagent formula, BC-5390 is an ideal solution that can better dierentiate white blood cells on anti-aging samples under room temperature
1 hour 4 hours 8 hours 12 hours 24 hours
Principles
Impedance method for WBC/BAS, RBC and PLT counting;
Cyanide free reagent for hemoglobin test;
Flow Cytometry (FCM) + Laser scatter + Chemical dye method for WBC difierential analysis
Parameters
21: WBC, LYM(#,%), NEU(#,%), MON (#,%), EOS (#,%), BAS (#,%), RBC, HGB,
HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV
3 histograms and 1 scattergram
Analysis Mode
Autoloader Whole Blood Mode
Closed Whole Blood Mode
Closed Predilute Mode
Throughput
Up to 60 samples per hour with autoloader
Up to 51 samples per hour with closed mode
Up to 53 samples per hour with predilute mode
Sample Volume
Whole Blood CBC+DIFF 33 uL
Whole Blood CBC 24 uL
Prediluted: 20 uL
Test Panel
CBC CBC+DIFF
Performance
Parameter |
Linearity Range |
Precision |
Carryover |
WBC (x 103/uL) | 0.3-200 |
≤0.15 (SD) or 3.0% (CV%) |
≤1.0% |
RBC (x 106/uL) | 0.2-8.0 |
≤1.5% |
≤1.0% |
HGB(g/dL) |
0.5-25 |
≤1.5% |
≤1.0% |
HCT (%) | 2 -75 | ≤2.0% | ≤1.0% |
PLT(x 103 /uL) |
5-2000 |
≤7.5 (SD) or 5.0% (CV%) |
≤1.0% |
Data Storage Capacity
Up to 100,000 results including numeric and
graphical information
Communication
LAN Port supports HL7 protocol
Operating Environment
Temperature: 15 °C~30°C
Humidity: 30~85%
Air Pressure: 70~106 kPa
Power Requirement
A.C.100-240V ≤ 300VA 50/60Hz
Dimension and Weight
Dimension (WxDxH, inches): 22.4 x 23.2 x 20.6
Weight: ≤ 143 pounds