Mindray BC-760 CS Auto Hematology Analyzer

BC-760 CS Auto Hematology Analyzer

Phone:+86-21-54286005

Resources

Auto Hematology Analyzer with ESR

BC-700 Series

Above and Beyond

 

BC-700 Series

SF Cube · Hybrid Channel PLT · Easy-W ESR
Above and Beyond

 

Above your expectations

SF Cube fluorescent technology allows reliable counting and diff­erentiation of abnormal samples

 

More refined and reliable cell differentiation

 

More reliable measurements for low-value samples

 

More comprehensive alarm messages for abnormalities

 

Beyond your expectations

Optical PLT-H in every CBC + DIFF test

 

In order to solve the above problem, we have developed a brand new parameter PLT-H.
It combines small PLTs from the conventional impedance method and large PLTs from the optical method. The solution can resist the interferences in conventional PLT detection without requiring extra reagents.

Schematic diagram of PLT-H

 

Easy-W ESR

 

CBC + DIFF + ESR in one test provide reliable ESR results with greater ease

The BC-700 series integrates an automatic ESR module in a hematology analyzer. It can also generate both CBC & ESR results in one test within 1.5 min. In addition, it saves the costs that would otherwise be incurred for the purchase, maintenance, consumables, and storage space of a separate ESR analyzer. Compared with the traditional Westergren method, this method performs better in quality traceability, repeatability, speed, safety, and level of automation.

 

Accurate

  • Great correlation with the Westergren method
  • Same QC and calibrator as in the BC-6000 series
  • Combined examination helps to avoid the interferences of dehydration, polycythemia vera and anemia on ESR results

 

Cost-effective

  • The integrated instrument is capable of both CBC and ESR detection
  • Takes up the space of only one analyzer

 

Automatic

  • Report CBC + ESR results together within 1.5 min
  • The measurement results are protected against the influence of subjective factors
  • Automation can reduce the biosafety hazards that may otherwise be introduced by a manual method

 

Excellent performance, high reliability, and ease of use

 

An all-in-one solution that goes above and beyond your expectations

Key Specifications

Principles
WBC (IMG/Neu/Mon/Lym/Eos/Bas), NRBC/RET*, PLT-H/PLT-O*/IPF:
SF Cube ^ Cell Analysis Technology
^S: Scatter; F: Fluorescence; Cube: 3D analysis

 

RBC, PLT
Focusing Flow-DC Impedance Method


HGB
Colorimetric method


ESR
Photometric method


Number of measuring parameters (whole blood): 109
Number of reportable parameters: 41
WBC Bas# Bas% Neu# Neu% Eos# Eos% Lym# Lym% Mon#
Mon% IMG# IMG% RET%* RET#* RHE* IRF* LFR* MFR* HFR*
RBC HGB MCV MCH MCHC RDW-CV RDW-SD HCT NRBC#
NRBC% PLT PLT-I PLT-H PLT-O* MPV PDW PCT P-LCR P-LCC
IPF ESR
Number of research parameters: 68*


Number of measuring parameters (body fluid): 18
Number of reportable parameters: 7
WBC-BF TC-BF# MN# MN% PMN# PMN% RBC-BF
Number of research parameters: 11


Sample volume
CD (whole blood): 25ul CD+ESR
(whole blood): 160ul Predilute: 20ul


Data storage capacity
Up to 150,000 results including numeric and graphical
information *


Throughput
CD 80t/h CDR 45t/h CD+ESR 40t/h


AnalysisMode

Sample Type

Analysis Mode

Whole blood

CBC, CBC + DIFF, CBC + DIFF+RET*, CD + ESR,CDR + ESR*, CD/WBC-3X, CDR/PLT-5X*,and other modes

Predilute

CBC, CBC + DIFF, CDR*, and other modes

Body fluid

CBC + DIFF

 

Physical Specifications

Dimensions
840D x 655W x 600H mm


Weight
≤70.6Kg


Voltage
100V-240V~ (±10%)


Frequency
50Hz/60Hz (±1Hz)


Power input
600VA


External output
LANx1 , USB x 4 (Specifications: DC 5V; 500mA;USB2.0 x 3; USB3.0 x 1 )

 

Normal Operating Environment

Ambient temperature:
10℃ ~ 35℃


Relative humidity:
30% ~ 85%


Atmospheric pressure:
70.0kPa ~ 106.0kPa^
^Note : Required altitude for normal operation:
-400m ~ +3000m

 

Performance

Parameter

Linearity Range

Precision

Carryover

WBC

0-500×109 /L

≤2.5% (≥4.51×109 /L)

≤ 1.0%

RBC

0-8.60×1012/L

≤1.5% (≥3.5×1012/L)

≤ 1.0%

HGB

0-260g/L

≤1.0% (110-180g/L)

≤ 1.0%

HCT

0-75%

≤1.5% (30%-50%)

≤ 1.0%

PLT*

0-5000×109 /L

≤ 1.5(SD) (≤20×109/L)^ ≤2.5% (≥100×109/L)^

≤ 1.0%

RET*

0-0.8×1012/L

≤15% (RBC ≥ 3.00×1012/L
RET%: 1.00% ~ 4.00%)

≤ 1.0%

ESR

 

≤1.8(SD)(0~20mm/h)

≤ 1.0%

Note: Applicable only to CDR/PLT-O 5x and CR/PLT-O 5x models

Items marked with an aerisk (*) apply only to BC-780