Mindray BC-30 Auto Hematology Analyzer
BC-30 Auto Hematology Analyzer
Phone:+86-21-54286005
Resources

Microsystem
Endoscopysystem
Energysystem
BC-30 Auto Hematology Analyzer
Phone:+86-21-54286005
Auto Hematology Analyzer
Minimum Size, Maximum Capability
Gives you confidence in results
Advanced detailed flag information provides warning in case of abnormal cell results. High quality original reagent calibrator and quality controls maintain your work at a high level. We integrate technologies into our products to fulfill
your clinical needs, therefore, today, Mindray products and services are used by healthcare facilities in over 190 countries and regions. More and more customers choose Mindray products and trust Mindray's quality.
Gives you less downtime
Quality and user experience are top priorities for Mindray. We design and manage our manufacturing processes in compliance with FDA, ISO and CE.Complete quality control systems in Mindray's manufacturing centers ensure we exceed the most stringent criteria and traceability throughout the entire process. Our professional and responsive support team makes sure your systems are kept at full capability.
Gives you more time to work
Smallest Convenience fast Perfect for pediatric samples
Smart and intuitive software bring your productivity to a new level. The 10.4 inch TFT touch screen makes operations easy and flexible, maximizing your productivity. The incredible 400,000 storage capacity gives you a worry-free solution to managing previous patients' data. The smart software can detect and remove operational errors, as well as run automatic maintenance with a single click.
Principles
Impedance method for WBC, RBC and PLT counting
Cyanide free reagent for hemoglobin test
Parameters
21 parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC,
HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR,P-LCC
3 histograms for WBC, RBC and PLT
Sample volume
Prediluted mode ≤20uL
Whole blood mode ≤9 uL
Performance
Parameter |
Linearity Range |
Precision |
Carryover |
WBC |
0-100×109/L |
≤3.5% (4-6.9×109/L) ≤2.0% (7.0-15.0×109/L) |
≤0.5% |
RBC |
0-8.00×1012/L |
≤1.5% (3.5-6.5×1012/L) |
≤0.5% |
HGB |
0-280g/L |
≤1.5% (100-180g/L) |
≤0.5% |
MCV(fL) | ≤1.0% (70.0-110.0) | ≤0.5% | |
PLT |
0-1000×109/L |
≤5.0% (100-149×109/L) ≤4.0% (150-500×109/L) |
≤1.0% |
Throughput
60 samples per hour
Display
10.4 inch TFT Touch Screen
Multi-language
Chinese, English, Spanish, Portuguese, Russian, French, Bahasa
Indonesia
Data Storage Capacity
Up to 100,000 results including numeric and graphical information
Communication
LAN Port supports HL7 protocol
Support bi-directional LIS
Interface
4 USB ports (for external printer, software upgrade, barcode reader,
external WIFI, keyboard and mouse), LAN port (1)
Printout
Thermal recorder, 50 mm wide paper, various printouts formats,External printer optional
Power Requirement
Temperature: 15°C~30°C
Humidity: 10%~90%
Air pressure: 70kPa~106kPa
Power Requirement
(100V-240V)±10%
≤180VA
(50Hz/60Hz) ±1Hz
Dimension and Weight
Dimension: Depth (410 mm) x width (300 mm) x height (400 mm)
Weight: ≤20Kg